可程式高低溫箱標準中規定的產品結構及外觀要求
The product structure and appearance may be required to program the GDW standards
經常瀏覽本站的朋友會發現,我們關于可程式高低溫箱標準中規定的相關要求,之前一直都是描述的標準起源與術語定義,以及設備的一些外部條件和內容環境等,今天,我們一起來了解下標準中對于產品結構與外觀的要求,也希望此篇文章能給正在對可程式高低溫箱型的廣大新老客戶們有一些小小的幫助,標準的就是的:高低溫試驗箱,高低溫交變試驗箱,高低溫交變濕熱試驗箱。
Often visit the site to friends will find, we related requirements may be required to program the GDW standards, until is to define the origin and terminology standard description, equipment and some external conditions and environment, today, we come to know the requirements for product structure and appearance of the standard, also hope this article can give is to programmable high-low temperature box type of the vast number of old and new customers with a little help, the standard is the authoritative: high and low temperature test, high temperature test box, alternating high and low temperature test chamber.
1.首先可程式高低溫箱內壁應使用耐熱不易氧化和具有一定機械強度的材料制造。應無影響試驗的污染源。
Manufacturing 1 first test box wall should use the heat is not easy to be oxidized and a certain mechanical strength of materials. Should not affect the pollution source test.
2.保溫材料應能耐高溫并具有阻燃性能。保溫層應有足夠的厚度,能保證試驗箱外部已觸及部位的溫度在高溫試驗時不高于50℃,在低溫試驗及環境溫度為15℃~35℃、相對濕度≤85%時不應有凝露現象。
2 insulation material should resist high temperature and flame retardant properties. Insulation shall be of sufficient thickness, can ensure the test box external has hit parts of the temperature is not higher than 50 ℃ in high temperature test, tests at low temperatures and environmental temperature is 15 ℃ ~35 ℃, relative humidity below 85%, there should be no condensation phenomenon.
3.加熱和制冷器件和冷量不應直接輻射在試驗樣品上。
3 heating and cooling device and cooling capacity should not direct radiation on test specimens.
4.工作室應設有觀察窗和照明裝置。
Studio 4 should be provided with an observation window and lighting device.
5.箱門應密封良好,密封條應有良好的抗高溫老化、耐低溫硬化性能。
The 5 door should be sealed well, seal shall have good resistance to high-temperature, low temperature resistance and hardening property.
6.制冷系統不應有漏氣、漏水、漏油缺陷。
6 refrigeration system should not leak, leakage, Water Leakage defects.
7.應有放置或懸掛試驗樣品的樣品架。樣品架應有足夠的耐高溫、低溫性能。
The sample holder should be placed or hung 7 test samples. The sample frame should have enough resistance to high temperature, low temperature performance.
8.應設有引線孔。
8 shall be provided with a pin hole.
9.外觀涂鍍層應平整光滑、色澤均勻,不得有露底、起泡、起層或擦傷痕跡。
9 the appearance of coating should be smooth, uniform color, not to betray a confidence, foaming, layer or scratches.
附:可程式高低溫箱安全和環境保護要求:
Attached: safety and environmental protection requirements:
(1).接線端子對箱體金屬外殼之間的絕緣電阻值應滿足:冷態2MΩ以上,熱態1MΩ以上(用500V,準確度為1.0級兆歐表測量),并能承受50Hz交流電壓1500V、施壓時間5s耐電壓試驗。
(1). The terminal shall meet the insulation resistance between the metal shell of the value of more than 2M Omega: cold, hot 1M Ω above (with 500V, the accuracy is 1 grade megger measurement), and can withstand 50Hz AC voltage 1500V, time 5S voltage withstanding test pressure
(2).保護接地端子應與試驗箱外殼有良好的電氣連接并能方便牢固的地接線,應符合GB/T14048.1—2000的7.1.9的規定。
(2). The protective earth terminal should have good electrical connection and convenient for firm connection and test case, shall be in accordance with the provisions of the 7.1.9 GB/T14048.1 2000.
(3).應有超溫、過電流、缺水等保護及報警裝置。
(3). And the alarm should be over temperature, over-current protection device, water etc..
(4).整機噪聲應不高于80dB(A)。
(4). The noise should be not more than 80dB (A).
立即詢價
您提交后,專屬客服將第一時間為您服務